WF Ultrapure Water System

RO pure water: desalination rate as high as 95-99% , Conductivity ≤10μS/cm
Ultrapure water:
resistivity 18.25MΩ.cm@25°C,
TOC: <3 ppb,
particles (>0.22 um) <1/ml,
microorganism <1 cfu/ml,
pyrogen<0.001EU/ml,
RNases≤0.01ng/ml,
DNases≤3Pg/ul

Feature
  • Applicable to various water quality, the case is with powder coating.
  • All filter elements in the whole system adopt open mold injection technology and built-in chassis
  • Korean quick-connect filter column is easy to replace and maintain.
  • The electrical and waterways are completely separated to avoid circuit failure caused by moisture and water leakage.
  • Water quality online monitoring system.
  • Automatic RO membrane anti-scaling flushing program, automatic shutdown of tap water, automatic water cut off, automatic water supply in the storage bucket, automatic shutdown of the water storage tank.
  • All pipelines are NSF certified and have water quality and water upgrade capabilities.
  • Equipped with a portable TDS pen, you can test TDS content, conductance, water temperature anytime, anywhere.
  • Double electric water intake, two water quality, ready to use.
  • Open the door design before and after, it is quick and easy to replace the filter element, and the filter replacement can be easily completed without using any tools.
Application

Applicable to all kinds of physical and chemical detection, biochemical analysis, blood examination, micro analysis, washing, HPLC, IC, GC, analytical experiments.

Process
Filter function
PP cotton filter remove impurities (greater than 5 micron) such as suspended solid, rust, sludge, colloid, etc.
Granular activated carbon filter remove color, odor, residual chlorine and heavy metals
Reverse osmosis membrane filter out <1nm pollutants, ions, particles
Ultra-purification column ion exchange resin bed, H+ and OHion removal
0.2μm terminal filter final removal for < 0.2μm bacteria
Specification
Model WF-UP WF-UP-F WF-UP-V WF-UP-VF
Type classification Standard Low Endotoxin Low TOC Comprehensive
Output 10/20/30 L/H
Water requirements Urban tap water, inlet water pressure: 0.15-0.5mpa, water temperature 5-40℃
RO Water Parameter
Ion rejection rate 97%-99% (when using a new RO membrane)
Organic matter rejection rate >99% when MW>200 Dalton
Particle and bacterial rejection > 99%
Application Front system of ultra-pure water machine, buffer configuration, photosensitive film washing, water supply for stability experiment box, meteorological experiment equipment, hydrogen generator, battery adding liquid, electroplating and other instruments
UP Water Parameter (ASTM TYPE I Grade Water)
Resistivity (25℃) 18.25 M Ω. Cm @ 25 ℃
Heavy metal ion < 0.01 ppb
Total organic carbon (TOC) * <10 ppb <3 ppb
bacteria < 1 cfu/ml * (with terminal filter)
Particulate matter (> 0.1 microns) <1/ml
Pyrogen (endotoxin) N/A < 0.001 Eu/ml N/A < 0.001 Eu/ml
Ribonuclease (RNases) N/A <1pg/ml N/A <1pg/ml
Deoxyribonucleases N/A <5pg/ml N/A <5pg/ml
Application All kinds of physical and chemical testing, biochemical analysis, blood examination, microanalysis, flushing special, HPLC, IC, GC, experiment High sensitivity ICP/MS, PCR, PPT level analysis, isotope analysis, molecular biology, life science, CDC, drug control institute, quality inspection institute Atomic absorption/emission spectrometry, mass spectrometry, TOC detection, environmental monitoring, biochip development, trace organic matter analysis Molecular biology, Life sciences, gene research, cell culture, amino acid analysis, protein purification, toxicology, drug development, medical testing
Electrical requirements 100-240V, 50/60Hz, 48-72W
The system configuration Host machine (including 1 set of purification column)+ 12L pressure water tank +TDS pen + accessory bag
Consumables
No. Filter element Replace Cycle Remark
1 11″ PP filter About 4-6 month
2 11″ activated carbon filter About 4-6 month
3 75 GPD RO membrane About 1-2 years
4 100 GPD RO membrane About 1-2 years
5 11″ ultra-purification column (imported resin) About 1000 liter
6 0.2 μm PES terminal filter About 2-3 years
7 Imported 5000 Dalton UF heat source column ——
8 Dual wavelength (185 & 254 nm) UV lamp (import) About9000 hours